Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 738

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
738User Manual 1802.3369.02 ─ 02
INRange | OORange
In range / out of range: Set the minimum and maximum value of
the range with TRIGger:LIN:IMIN and TRIGger:LIN:IMAX.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 369
TRIGger:LIN:IMAX <IdPattern>
Sets the end value of an identifier range if TRIGger:LIN:ICONdition is set to
INRange or OORange.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 369
TRIGger:LIN:IMIN <IdPattern>
Specifies a secondary identifier pattern, or sets the start value of an identifier range.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 369
TRIGger:LIN:IPERror <IdParityError>
Triggers if an error occurs in the identifier parity bits. These are the bits 6 and 7 of the
identifier.
Available, if TRIGger:LIN:TYPE is set to ERRC.
Parameters:
<IdParityError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 370
TRIGger:LIN:SYERror <SyncError>
Triggers if a synchronization error occurs.