Remote control commands
R&S
®
MXO 5 Series
481User Manual 1802.3369.02 ─ 02
17.8.8.2 Settings for active voltage probes
The probe suffix <ch> selects the input channel to which the probe is connected.
PROBe<ch>:SETup:MODE.............................................................................................481
PROBe<ch>:SETup:ACCoupling..................................................................................... 482
PROBe<ch>:SETup:CMOFfset........................................................................................482
PROBe<ch>:SETup:DISPlaydiff.......................................................................................483
PROBe<ch>:SETup:ZAXV.............................................................................................. 483
PROBe<ch>:PMETer:STATe............................................................................................483
PROBe<ch>:PMETer:RESults:SINGle?............................................................................484
PROBe<ch>:PMETer:RESults:COMMon?........................................................................ 484
PROBe<ch>:PMETer:RESults:DIFFerential?.....................................................................484
PROBe<ch>:PMETer:RESults:NEGative?........................................................................ 485
PROBe<ch>:PMETer:RESults:POSitive?..........................................................................485
PROBe<ch>:SETup:ADVanced:AUDioverload.................................................................. 486
PROBe<ch>:SETup:ADVanced:FILTer............................................................................. 486
PROBe<ch>:SETup:ADVanced:RANGe........................................................................... 486
PROBe<ch>:SETup:ADVanced:PMToffset........................................................................ 487
PROBe<ch>:SETup:MODE <Mode>
The micro button is located on the probe head. Pressing this button, you initiate an
action on the instrument directly from the probe. The button is disabled during internal
automatic processes, for example, during self-alignment, autoset, and level detection.
Select the action that you want to start from the probe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Mode> RCONtinuous | RSINgle | AUToset | AZERo | OTMean | SITFile |
NOACtion | FTRiglevel | PRSetup
RCONtinuous
Run continuous: the acquisition is running as long as the probe
button is pressed.
RSINgle
Run single: starts a defined number of acquisitions (same as
[Single] key).
AUToset
Starts the autoset procedure.
AZERo
AutoZero: performs an automatic correction of the zero error.
OTMean
Set offset to mean: performs an automatic compensation for a
DC component of the input signal.
SITFile
Save image to file: saves the display image in a file.