Instrument setup
R&S
®
MXO 5 Series
91User Manual 1802.3369.02 ─ 02
5.6.4 Restoring settings
When you have changed many settings on the instrument, and you are not sure which
settings are causing which effect, you can restore the default settings and start anew.
The following methods are available:
Saving instrument settings as a user-defined preset and restoring the instrument
settings to user-defined default values
Restoring all settings on the R&S MXO 5 to the factory-defined values
Restoring settings from a file
To save a user-defined preset
1. Open the "Menu" > "Settings" > "Preset" tab.
2. Enter a name for the preset file. Select the file format.
3. Tap "Save preset file".
Note: If you want to store the file in another directory than the displayed one,
select "Save as". See also: Chapter 12.5, "File browser dialog", on page 283.
To restore the instrument settings to user-defined default values
1. Open the "Menu" > "Settings" > "Preset" > "Settings" tab.
2. In "Preset file name", enter the name of the file that contains the required settings.
3. To use these settings as preset values, select "Enable user-defined preset".
4. Press the [Preset] key.
To restore all settings to the factory defaults
1. Open the "Menu" > "Settings" > "Preset" > "Factory" tab.
2. Tap the "Factory preset" button.
All settings on the R&S MXO 5 are reset to their factory-defined values.
As long as no user-defined preset file is loaded and "Enable user-defined preset" is
disabled, the [Preset] key also resets the instrument settings to factory defaults.
5.7 Maintenance settings
In the "Settings" > "Maintenance" dialog box, you can update the firmware, perform
self-alignment.
5.7.1 Firmware update
Access: "Menu" > "Settings" >"Maintenance" > "FW update".
Maintenance settings