Remote control commands
R&S
®
MXO 5 Series
727User Manual 1802.3369.02 ─ 02
<fr> index of the frame
Return values:
<FrameState> OK | EOFD | ACKD | NOACk | CRCD | CRC | SERRror |
FORM | BTST | UNKNown | INComplete
OK: the frame is valid.
EOFD: Wrong end of frame.
ACKD: Wrong ACK delimiter occurred.
NOACK: Acknowledge is missing.
CRCD: Wrong CRC delimiter occurred.
CRC: Cyclic redundancy check failed.
SERRror: Stuff count error (CAN FD ISO only).
FORM: Fixed-bit form error.
BTST: Bit stuffing error occurred.
INSufficient: The frame is not completely contained in the
acquisition. The acquired part of the frame is valid.
*RST:
OK
Usage: Query only
Asynchronous command
SBUS<sb>:CAN:FRAMe<fr>:STOP?
Return the stop time of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameStop> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:CAN:FRAMe<fr>:STUFf?
Returns the value of the stuff count field.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<StuffCount> Range: 0 to 7
Increment: 1