Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 175

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Waveform analysis
R&S
®
MXO 5 Series
175User Manual 1802.3369.02 ─ 02
To display a particular acquisition
You can analyze a selected history waveform, and save it to file.
In the quick-access "History" dialog box, enter the number of the required acquisi-
tion in the "Current acq" field. The newest acquisition always has the index "0",
older acquisitions have a negative index.
Alternatively, you can configure and start the history display from the "History" configu-
ration dialog box:
1. Open the "History" configuration tab.
2. If the history mode is off (the [History] key is not illuminated), select "Show history".
3. Drag the slider to the required acquisition. The current number is shown in the
"Current acq" field.
Alternatively, enter the number of the required acquisition in the "Current acq" field.
To save a single history waveform
1. Select and display the required history waveform as described in "To display a par-
ticular acquisition" on page 175.
2. Save the selected waveform using "Save/Recall" > "Save" tab > "Waveform". For
details, see Chapter 12.2, "Save and recall waveform data", on page 271.
To replay history waveforms
If you want to see the complete acquisition series without any setup, simply tap in the
quick-access "History" dialog box. For specific analysis of history data, use the history
"Viewer" setup.
1. Open the "History" configuration dialog tab.
2. If the history mode is off (the [History] key is not illuminated), enable "Show his-
tory".
3.
Tap
to start.
To exit the history
Choose one of the following ways:
Close the quick-access "History" dialog box.
In the "History" configuration tab, disable "Show history".
History
Loading ...
Loading ...
Loading ...