Remote control commands
R&S
®
MXO 5 Series
485User Manual 1802.3369.02 ─ 02
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Result> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Manual operation: See "ProbeMeter" on page 124
PROBe<ch>:PMETer:RESults:NEGative?
Returns the R&S ProbeMeter measurement result of differential active R&S probes,
the voltage that is measured between the negative signal socket and the ground.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Result> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Manual operation: See "ProbeMeter" on page 124
PROBe<ch>:PMETer:RESults:POSitive?
Returns the R&S ProbeMeter measurement result of differential active R&S probes:
the voltage that is measured between the positive signal socket and the ground.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Result> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Manual operation: See "ProbeMeter" on page 124