Remote control commands
R&S
®
MXO 5 Series
759User Manual 1802.3369.02 ─ 02
<n> 1..4
Selects the channel group:
1 = digital channels 0..3;
2 = digital channels 4..7
3 = digital channels 8..11
4 = digital channels 12..15
Parameters:
<Threshold>
Usage: Asynchronous command
Manual operation: See "Technology,Threshold" on page 380
17.17.3
MSO data
The remote data transfer from the instrument to the controlling computer is performed
using the following commands:
DIGital<m>:DATA:HEADer?............................................................................................ 759
DIGital<m>:DATA[:VALues]?........................................................................................... 759
PBUS<pb>:DATA:FORMat..............................................................................................760
PBUS<pb>:DATA:HEADer?............................................................................................ 760
PBUS<pb>:DATA[:VALues]?........................................................................................... 761
DIGital<m>:DATA:HEADer?
Returns the header of digital channel data
Table 17-7: Header data
Position Meaning Example
1 XStart, acquisition time before trigger, in s -5E-008 = - 50 ns
2 XStop, acquisition time after trigger, in s 5E-008 = 50 ns
3 Record length of the waveform in Samples 1000
4 Number of values per sample interval. For digital
data, the result is 1.
1
Suffix:
<m>
.
0..15
Selects the digital channel.
Usage: Query only
Asynchronous command
DIGital<m>:DATA[:VALues]? [<Offset>],[<Length>]
Returns the data of the indicated digital channel for transmission from the instrument to
the controlling computer. The data can be used in MATLAB, for example.
Mixed signal option (option R&S