Remote control commands
R&S
®
MXO 5 Series
483User Manual 1802.3369.02 ─ 02
PROBe<ch>:SETup:DISPlaydiff <DisplayDiff>
Selects the voltage to be measured by the R&S ProbeMeter of differential active
probes:
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<DisplayDiff> DIFFerential | SINGleended
DIFFerential
Measures differential and common mode voltages
SINGleended
Measures the voltage between the positive/negative signal
socket and the ground.
*RST: DIFFerential
Usage: Asynchronous command
Manual operation: See "Display" on page 126
PROBe<ch>:SETup:ZAXV <AttenuationZA15>
If you use the external attenuator R&S RT-ZA15 together with one of the differential
active probes R&S RT-ZD10/20/30, enable "RT-ZA15 attenuator" to include the exter-
nal attenuation in the measurements.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<AttenuationZA15> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "RT-ZA15 attenuator" on page 126
PROBe<ch>:PMETer:STATe <State>
Activates the integrated R&S ProbeMeter on probes with Rohde & Schwarz probe
interface.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<State> ON | OFF
*RST: OFF
Acquisition and setup