Remote control commands
R&S
®
MXO 5 Series
559User Manual 1802.3369.02 ─ 02
Parameters:
<Source> NONE | C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 |
D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 |
M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8
Usage: Asynchronous command
MEASurement<mg>:SSRC <Source2>
Defines the second measurement source.
The command is an alternative to MEASurement<mg>:SOURce.
Suffix:
<mg>
.
1...24, index of the measurement
Parameters:
<Source2> NONE | C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 |
D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 |
M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8
Usage: Asynchronous command
Manual operation: See "Clock source" on page 209
MEASurement<mg>:MAIN <MeasType>
Defines the measurement that is used as a source for math calculations and result
analysis.
Suffix:
<mg>
.
1...24, index of the measurement
Parameters:
<MeasType> HIGH | LOW | AMPLitude | MAXimum | MINimum | PDELta |
MEAN | RMS | STDDev | CRESt | POVershoot | NOVershoot |
AREA | RTIMe | FTIMe | PPULse | NPULse | PERiod |
FREQuency | PDCYcle | NDCYcle | CYCarea | CYCMean |
CYCRms | CYCStddev | CMAXimum | CMINimum | CPDelta |
PULCnt | DELay | PHASe | BWIDth | EDGecount | SETup |
HOLD | SHT | SHR | DTOTrigger | SLERising | SLEFalling
Usage: Asynchronous command
MEASurement<mg>:ENVSelect <EnvelopeCurve>
Relevant only for measurements on envelope waveforms. It selects the envelope to be
used for measurement.
Prerequisites:
ACQuire:TYPE on page 475 is set to ENVElope.
Automatic measurements
Loading ...
Loading ...
Loading ...