![](https://files.manualsfile.com/11202903-mxo54-2000/bg1e6.png)
Remote control commands
R&S
®
MXO 5 Series
486User Manual 1802.3369.02 ─ 02
PROBe<ch>:SETup:ADVanced:AUDioverload <Sound>
Activates the acoustic overrange warning in the probe control box. The command is
relevant for R&S RT-ZHD probes.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Sound> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Audible overrange" on page 129
PROBe<ch>:SETup:ADVanced:FILTer <State>
Activates the lowpass filter in the probe control box. The filter frequency depends on
the probe type and is indicated on the probe control box.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Bandwidth limit" on page 129
PROBe<ch>:SETup:ADVanced:RANGe <ProbeRange>
Sets the voltage range of an R&S RT-ZHD probe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<ProbeRange> AUTO | MHIGh | MLOW
AUTO
The voltage range is set with CHANnel<ch>:SCALe.
MHIGh
Sets the higher voltage range of the connected probe. To query
the value, use PROBe<ch>:SETup:ATTenuation[:AUTO]?.
MLOW
Sets the lower voltage range of the connected probe. To query
the value, use PROBe<ch>:SETup:ATTenuation[:AUTO]?.