Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 563

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
563User Manual 1802.3369.02 ─ 02
MEASurement<mg>:GATE <Gate>
Sets the gate of the indicated measurement. Enable a gate before you assign a mea-
surement to it (GATE<g>:ENABle =ON).
The query returns 0, if no gate is assigned.
Suffix:
<mg>
.
1...24, index of the measurement
Parameters:
<Gate> Number of the gate to be used
Usage: Asynchronous command
Manual operation: See "Measurements gated by G<x>" on page 198
MEASurement<mg>:REFLevel<sc>:REFerence
<RefLevelSet>
Selects the set of reference levels that is used for the measurement and for the indica-
ted source. Each source of the measurement can have its own reference level set.
Suffix:
<mg>
.
1...24, index of the measurement
<sc> 1...3
Source waveform, to which the reference level set is applied.
Parameters:
<RefLevelSet> Number of the reference level set. Define the reference level set
before you use it.
Example:
MEASurement1:MAIN DELay
MEASurement1:SOURce C1,C2
MEASurement1:REFLevel1:REFerence 1
MEASurement1:REFLevel2:REFerence 2
Measurement 1 is a delay measurement on channel 1 and chan-
nel 2. C1 is the first source and uses reference level set 1 (RL1).
C2 is the second source and uses reference level set 2 (RL2).
Usage: Asynchronous command
Manual operation: See "Reference levels" on page 207
17.12.3 Results
MEASurement<mg>:RESult[:ACTual]?.............................................................................564
MEASurement<mg>:RESult:AVG?...................................................................................564
MEASurement<mg>:RESult:NPEak?............................................................................... 564
MEASurement<mg>:RESult:PPEak?............................................................................... 564
MEASurement<mg>:RESult:RELiability?..........................................................................564
MEASurement<mg>:RESult:RMS?.................................................................................. 564
MEASurement<mg>:RESult:WFMCount?.........................................................................564
Automatic measurements
Loading ...
Loading ...
Loading ...