Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
789User Manual 1802.3369.02 ─ 02
STATus:QUEStionable:LIMit:CONDition?
MEAS1,MEAS2
// Measurements 1 and 2 are still exceeding the limit
STATus:QUEStionable:EVENt?
0
// Waiting for an event
*SRQ?
// Measurement 3 exceeds the limit
STATus:QUEStionable:LIMit:CONDition?
MEAS1,MEAS2,MEAS3
// Measurements 1,2 and 3 are currently exceeding the limit
STATus:QUEStionable:LIMit:EVENt?
MEAS3
// Measurement 3 has newly exceeded the limit
// The event register is cleared after the query
Example: Querying with STATus:OPERation:CONDition
The following example shows how the result of STATus:OPERation:CONDition?
changes during a single acquisition.
SINGLe
// Measurement starts
STATus:OPERation:CONDition?
16
//Measuring
STATus:OPERation:CONDition?
48
//Pre trigger waiting time is over (16 Measuring + 32 WaitForTrigger)
STATus:OPERation:CONDition?
112
//Trigger is deactivated (16 Measuring + 32 WaitForTrigger + 64 Triggered)
STATus:OPERation:CONDition?
96
// Measurement has finished (32 WaitForTrigger + 64 Triggered)
Query OPC status in the event status register
If you activate a time-consuming operation and wait for completion with *OPC?, a time-
out could occur before the operation is finished and you do not receive the returned
Status reporting