Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 672

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
672User Manual 1802.3369.02 ─ 02
TRIGger:I2C:AMODe <AddressType>
Sets the address length to be triggered on: 7 bit or 10 bit.
Parameters:
<AddressType> BIT7 | BIT10
*RST: BIT7
Usage: Asynchronous command
Manual operation: See "Address type" on page 319
TRIGger:I2C:DCONdition
<DataOperator>
Sets the operator to set a specific data value or a data range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Data" on page 320
TRIGger:I2C:DMIN <Data>
Specifies the data bit pattern, or sets the start value of a data pattern range. Enter the
bytes in MSB first bit order. The maximum pattern length is 64 bit. Waveform data is
compared with the pattern byte-by-byte.
Parameters:
<Data>
Usage: Asynchronous command
Manual operation: See "Data" on page 320
TRIGger:I2C:DPOSition <DataPosition>
Sets the number of data bytes to be skipped after the address.
Parameters:
<DataPosition> Range: 1 to 4096
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Position" on page 320
TRIGger:I2C:DRNack <DataReadNack>
Triggers on the end of the read process when the controller reads data from the target.
This NACK is sent according to the protocol definition, it is not an error.
Protocols