Remote control commands
R&S
®
MXO 4 Series
722User Manual 1335.5337.02 ─ 05
Bits Source values Description
TEMPerature TEMPWARN
Temperature warning: indicates that the device is
getting hot and should be cooled.
TEMPERROR
Temperature error: the device is too hot and will
automatically shut down in a few seconds.
ADCState CPCHAN<m>
Positive clipping for analog channels
CNCHAN<m>
Negative clipping for analog channels
CPPRobe<m>
Positive clipping for probes
CNPRobe<m>
Negative clipping for probes
GOVerload WGENerator<m>
Available generators
NOALigndata HCHannel<m>
Horizontal channels
VCHannel<m>
Vertical channels
PROBe<m>
Probes channels
LPRObes<m>
Logic probes
WGENerator<m>
Available generators
LIMit MEAS<m>
Enabled measurements
MARGin MEAS<m>
Enabled measurements
IMPRecise MEAS<m>
Enabled measurements
Some of the status commands can be used interchangeably:
Table 17-9: Overview: STATus:QUEStionable and respective CHANnel, MEASurement and WGENera-
tor commands
STATus:QUEStionable register
MEASurement
STATus:QUEStionable:COVerload
Bit: CHANnel<ch>
CHANnel<ch>:OVERload:STATus
STATus:QUEStionable:COVerload
Bit: WCHannel<ch>
CHANnel<ch>:WARNoverload:STATus
STATus:QUEStionable:COVerload
Bit: EXTTRIGGERIN
TRIGger:ANEDge:OVERload:STATus
STATus:QUEStionable:COVerload
Bit: TRIGGEROUT
TRIGger:ACTions:OUT:OVERload:STATus
STATus:QUEStionable:ADCState
Bit: CPCHannel<ch>
CHANnel<ch>:ADCState:STATus:PCLipping
STATus:QUEStionable:ADCState
Bit: CNCHannel<ch>
CHANnel<ch>:ADCState:STATus:NCLipping
STATus:QUEStionable:PPSupply
Bit: CNCHannel<ch>
CHANnel<ch>PPSupply:STATus
Status reporting
Loading ...