Remote control commands
R&S
®
MXO 4 Series
702User Manual 1335.5337.02 ─ 05
Usage: Asynchronous command
Manual operation: See "Pulse width" on page 364
WGENerator<wg>:FUNCtion:RAMP[:SYMMetry] <RampSymmetry>
Sets the symmetry of a ramp waveform, the percentage of time the waveform is rising.
By changing the symmetry of the ramp, you can create, for example, triangular wave-
forms.
Suffix:
<wg>
.
1..2
Parameters:
<RampSymmetry> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Symmetry" on page 363
WGENerator<wg>:FUNCtion[:SELect] <FunctionType>
Selects the type of waveform to be generated.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<FunctionType> SINusoid | SQUare | RAMP | DC | PULSe | SINC | CARDiac |
GAUSs | LORNtz | EXPRise | EXPFall | ARBitrary
SINC: Cardial sine
*RST: SINusoid
Usage: Asynchronous command
Manual operation: See "Function type" on page 357
WGENerator<wg>:FUNCtion[:SQUare]:DCYCle <SquareDutyCycle>
Sets the duty cycle for the pulse function.
Suffix:
<wg>
.
1..2
Parameters:
<SquareDutyCycle> Range: 0.01 to 99.99
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command