Remote control commands
R&S
®
MXO 4 Series
520User Manual 1335.5337.02 ─ 05
MMEMory:RCL <FileSource>
Restores the instrument settings from the specified file.
The stored instrument settings do not include waveform generator settings.
This command has the same effect as the combination of MMEMory:LOAD:STATe and
*RCL.
Parameters:
'<FileSource>' String parameter specifying the path and filename of the source
file. Wildcards are not allowed.
Usage: Event
MMEMory:STORe:STATe <MemoryNumber>, <FileName>
Stores the instrument settings from the specified internal memory to the specified file.
To store the current instrument settings to the internal memory, use *SAV first.
Setting parameters:
<MemoryNumber> Number of the internal memory
Range: 1 to 99
<FileName> String parameter specifying the complete path and filename of
the source file.
Usage: Setting only
MMEMory:LOAD:STATe <MemoryNumber>, <FileName>
Loads the instrument settings from the specified file to the specified internal memory.
After the file has been loaded, the settings must be activated using a *RCL command.
Setting parameters:
<MemoryNumber> Number of the internal memory
Range: 1 to 99
<FileName> String parameter specifying the complete path and filename of
the source file.
Usage: Setting only
MMEMory:GENerator:RCL <FilePath>
Restores the waveform generator settings from the specified file.
Parameters:
<FilePath> String parameter specifying the path and filename of the settings
file.
Usage: Asynchronous command
Data management