Acquisition and waveform setup
R&S
®
MXO 4 Series
105User Manual 1335.5337.02 ─ 05
"Manual"
The sample rate is manually defined with "Sample rate", "Min. sample
rate".
Remote command:
ACQuire:SRATe:MODE on page 440
RL mode
Selects the mode of the waveform record length adjustment.
"Auto"
Record length is determined automatically and changes due to instru-
ment internal adjustments due to other setting changes.
"Manual"
The waveform record length is manually defined with "Record length",
"Record length limit".
Remote command:
ACQuire:POINts:MODE on page 438
Sample rate, Min. sample rate
Sets the number of waveform points per second in manual sample rate mode. In auto-
matic sample rate mode, it sets the minimum value of the sample rate for automatic
calculation.
The sample rate considers the samples of the ADC, and the processing of the cap-
tured samples including interpolation.
The setting is relevant, if "SR mode" is set to "Manual".
Remote command:
ACQuire:SRATe[:VALue] on page 439
Record length, Record length limit
Sets the record length in manual record length mode. In automatic record length mode,
it sets the maximum value of the record length for automatic calculation.
The record length is the number of waveform samples that are stored in one waveform
record after processing, including interpolation.
Remote command:
ACQuire:POINts[:VALue] on page 438
N-single/Avg count
The acquisition and average count has several effects:
It sets the number of waveforms acquired with [Single].
It defines the number of waveforms used to calculate the average waveform.
Thus, the instrument acquires sufficient waveforms to calculate the correct average
if "Average" is enabled for waveform arithmetic. The higher the value is, the better
the noise is reduced.
It sets the number of acquisitions to be acquired in a fast segmentation acquisition
series, and in the history.
The setting is read-only if "Segmented" > "Acquire maximum" is "On". See also
"Acquire maximum" on page 107.
Remote command:
ACQuire:COUNt on page 437
Acquisition