Remote control commands
R&S
®
MXO 4 Series
487User Manual 1335.5337.02 ─ 05
Parameters:
<Relative> Range: 0 to 50
Increment: 1
*RST: 0
Default unit: %
Usage: Asynchronous command
Manual operation: See "Relative hysteresis" on page 164
TRIGger:NOISe<m>[:STATe] <Mode>
Selects how the hysteresis is set.
Suffix:
<m>
.
1 to 4, index of the analog channel
Parameters:
<Mode> AUTO | MANual
AUTO
Automatic mode is the recommended mode. The hysteresis is
set by the instrument to reject the internal noise of the instru-
ment.
MANual
The hysteresis is defined with TRIGger:NOISe<m>:ABSolute
or TRIGger:NOISe<m>:RELative.
Usage: Asynchronous command
17.9.16 Channel filter
TRIGger:FILTermode......................................................................................................487
TRIGger:LFReject..........................................................................................................487
TRIGger:RFReject......................................................................................................... 488
TRIGger:FILTermode <Mode>
Selects the filter mode for the trigger channel.
Parameters:
<Mode> OFF | LFReject | RFReject
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Trigger filter mode" on page 165
TRIGger:LFReject <Bandwidth>
Sets the limit frequency limit for the high-pass filter of the trigger signal. Frequencies
lower than this value are rejected, higher frequencies pass the filter.