Remote control commands
R&S
®
MXO 4 Series
650User Manual 1335.5337.02 ─ 05
Parameters:
<DLCOperator> EQUal | GETHan
For little endian transfer direction, EQUal must be set.
*RST: GETHan
Usage: Asynchronous command
Manual operation: See "DLC" on page 331
TRIGger:CAN:FTYPe <FrameType>
Sets the CAN frame type.
Parameters:
<FrameType> CBFF | CBFR | CEFF | CEFR | FBFF | FEFF | XLFF | ERRor |
OVERload
CBFF: classical base frame format data
CEFF: classical extended frame format data
FBFF: FD base frame format
FEFF: FD extended frame format
XLFF: XL frame format
ERR: error
OVLD: overload
*RST: CBFF
Usage: Asynchronous command
Manual operation: See "Frame type" on page 329
TRIGger:CAN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one identifier
pattern to be set with TRIGger:CAN:IMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger:CAN:IMIN and TRIGger:CAN:IMAX.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Identifier" on page 331