Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 279

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Protocol analysis
R&S
®
MXO 4 Series
279User Manual 1335.5337.02 ─ 05
Remote command:
SBUS<sb>:RESult on page 589
Decode layer
Selects the decode layer.
Decoding is performed in several steps, and the end results are presented in the
decode table. The decode layer selects an interim step for which the decoding result is
shown in the honeycomb display.
The available values are protocol-specific.
"Off"
No decode layer is displayed.
"Edges"
All edges.
"Bits"
All bits.
"Words"
All words.
Data format
Sets the data format for the values displayed in the decode table and in the combs of
the decoded signal.
Available formats are: hex, octal, binary, ASCii, signed, unsigned.
For more information on the available formats, see Chapter 13.1.8, "Bit pattern editor",
on page 282.
Remote command:
SBUS<sb>:FORMat on page 590
Zoom coupling
If enabled, the decode zoom and result table are synchronized. If you select a row in
the result table, this result is shown in the decode zoom.
Remote command:
SBUS<m>:ZCOupling on page 590
Symbols
Symbol lists are protocol-specific. They are described in the respective configuration
chapter of the protocol.
13.1.4 Filter
In the "Filter" tab, you can define the settings to display only the frames that match the
selected filter conditions.
The settings are protocol specific and are described in the corresponding protocol
chapter.
13.1.5 Trigger
The "Trigger" tab gives quick access to the "Trigger" dialog.
Basics of protocol analysis
Loading ...
Loading ...
Loading ...