Remote control commands
R&S
®
MXO 4 Series
468User Manual 1335.5337.02 ─ 05
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
Triggers on runts longer than the given runt width.
SHORter
Triggers on runts shorter than the given runt width.
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by runt width and ±Delta.
OUTSide
Triggers if the runt length is outside a given time range. The
range is defined by runt width and ±Delta.
*RST:
ANY
Usage: Asynchronous command
Manual operation: See "Range" on page 146
TRIGger:EVENt<ev>:RUNT:WIDTh <Width>
Defines the upper or lower voltage threshold.
It is not available if TRIGger:EVENt<ev>:RUNT:RANGe is set to ANY.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Width> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 5E-09
Default unit: s
Usage: Asynchronous command
Manual operation: See "Runt width" on page 146
17.9.7 Window trigger
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer.............................................................. 469
TRIGger:EVENt<ev>:LEVel<n>:WINDow:UPPer............................................................... 469
TRIGger:EVENt<ev>:WINDow:DELTa.............................................................................. 469
TRIGger:EVENt<ev>:WINDow:RANGe............................................................................ 470
TRIGger:EVENt<ev>:WINDow:TIME................................................................................470
TRIGger:EVENt<ev>:WINDow:WIDTh............................................................................. 471
Trigger
Loading ...
Loading ...