Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 148

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Trigger
R&S
®
MXO 4 Series
148User Manual 1335.5337.02 ─ 05
Lower level
Sets the lower voltage limit.
Remote command:
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer on page 469
Time condition
Available for "Vertical condition" = "Stay within"/"Stay outside".
Selects how the time limit of the window is defined.
"Within"
Triggers if the signal stays inside or outside the vertical window limits
at least for the time Width - Delta and for Width + Delta at the most.
"Outside"
"Outside" is the opposite definition of "Within". The instrument triggers
if the signal stays inside or outside the vertical window limits for a
time shorter than Width - Delta or longer than Width + Delta.
"Shorter"
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
"Longer"
Triggers if the signal crosses vertical limits after the specified "Width"
time is reached.
Remote command:
TRIGger:EVENt<ev>:WINDow:TIME on page 470
Width
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Remote command:
TRIGger:EVENt<ev>:WINDow:WIDTh on page 471
±Delta
Defines a range around the width value.
Remote command:
TRIGger:EVENt<ev>:WINDow:DELTa on page 469
7.4.7 Timeout trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Timeout"
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source does not
have the expected transition within the specified time.
Trigger types
Loading ...
Loading ...
Loading ...