Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 348

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Mixed signal option (MSO, R&S MXO4-B1)
R&S
®
MXO 4 Series
348User Manual 1335.5337.02 ─ 05
14 Mixed signal option (MSO, R&S MXO4-B1)
The Mixed Signal Option R&S MXO4-B1 adds logic analyzer functions to the classical
oscilloscope functions. Using the MSO option, you can analyze and debug embedded
systems with mixed-signal designs that use analog signals and correlated digital sig-
nals simultaneously.
The Mixed Signal Option provides 16 digital channels grouped in two logic probes
(pods) with 8 channels each. The instrument ensures that analog and digital wave-
forms are time-aligned and synchronized so that critical timing interactions between
analog and digital signals can be displayed and tested. The automatic alignment com-
pensates the skew between the probe connectors of the analog channels and the
probe boxes of the digital channels.
14.1 Logic configuration
Access: "Menu" > "Logic" > "Setup" tab
Digital channels can be displayed individually, and they can be grouped and displayed
as a logic group. You can configure and enable up to 4 logic groups. Each digital chan-
nel can be assigned to one active logic only, the instrument disables conflicting buses
automatically.
For clocked buses, you can display the decoded data in a result box.
If you have configured several logic groups and you want to modify the settings, make
sure that the tab of the correct logic is selected. Disable the logic before you change
the settings.
14.1.1 Setup
Access: "Menu" > "Logic" > "Setup" tab
Logic configuration
Loading ...
Loading ...
Loading ...