248User Manual 1335.5337.02 ─ 05
Maximum phase
Sets the upper boundary of the vertical phase window.
The lower boundary is given by "Maximum phase" - 360°.
By default, the "Maximum phase" is set to 180° for a phase window ranging from -180°
to 180° accordingly.
Remote command:
FRANalysis:PHASe:MAXimum on page 578
Measurement delay, Delay time
Sets a time delay, that the system waits before measuring the next point of the plot.
Time delay is helpful in systems that need more time to adapt to the new frequency, for
example if filters with significant time group delays are present.
Remote command:
FRANalysis:MEASurement:DELay:STATe on page 576
FRANalysis:MEASurement:DELay[:TIME] on page 576
RBW
Sets the resolution bandwidth, which determines the number of measurements that are
used for creating the plot.
"High (fast)"
A high RBW value is useful for obtaining a general understanding of a
system's behavior over a wide frequency range. The accuracy of the
measurement is not as good, but the measurement is completed fast.
"Medium"
A medium RBW value is a compromise between the measurement
time and measurement accuracy.
"Low (slow)"
A low RBW value allows for a more detailed analysis of the system.
Since more data needs to be collected, the measurement time would
increase significantly.
Remote command:
FRANalysis:MEASurement:RBW on page 577
Frequency response analysis (option R&S
MXO4-K36)
Loading ...
Loading ...