Trigger
R&S
®
MXO 4 Series
138User Manual 1335.5337.02 ─ 05
Remote command:
TRIGger:MEVents:SEQuence<se>:COUNt on page 459
Enable reset event
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the R-trigger conditions are fulfilled.
Remote command:
TRIGger:MEVents:SEQuence<se>:RESet:EVENt on page 460
Enable reset by time, Reset timeout
If reset timeout is enabled, the instrument waits for the "Reset timeout" time for the
specified number of B-triggers. If no trigger occurs during that time, the sequence is
restarted with the A-trigger.
Remote command:
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout[:ENABle] on page 461
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout:TIME on page 460
7.4 Trigger types
Edge trigger...........................................................................................................138
Edge trigger on external trigger source.................................................................139
Glitch trigger..........................................................................................................141
Width trigger..........................................................................................................143
Runt trigger........................................................................................................... 144
Window trigger...................................................................................................... 146
Timeout trigger...................................................................................................... 148
Interval trigger....................................................................................................... 149
Slew rate trigger....................................................................................................151
Setup & Hold.........................................................................................................153
State trigger...........................................................................................................155
Pattern trigger....................................................................................................... 156
Line trigger............................................................................................................ 161
7.4.1 Edge trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Edge"
The edge trigger is the most common trigger type. The trigger occurs when the signal
from the trigger source passes the trigger level in the specified direction (slope).
Trigger types
Loading ...
Loading ...