Remote control commands
R&S
®
MXO 4 Series
652User Manual 1335.5337.02 ─ 05
Parameters:
<FormError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 333
TRIGger:CAN:FDATa:SCERror <StuffCntErr>
Triggers on stuff count errors. A stuff bit error occurs, if more than five consecutive bits
of the same level occur on the bus.
Available, if TRIGger:CAN:TYPE is set to ERRor.
Parameters:
<StuffCntErr> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 333
TRIGger:CAN:ACKerror <AckError>
Triggers when the transmitter does not receive an acknowledgment - a dominant bit
during the Ack Slot.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<AckError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 333
TRIGger:CAN:BITSterror <BitStuffError>
Triggers if a stuff error occurs - when the 6th consecutive equal bit level in the men-
tioned fields is detected.
Available, if TRIGger:CAN:TYPE is set to ERRC.
Parameters:
<BitStuffError> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error conditions" on page 333