Network operation and remote control
R&S
®
MXO 4 Series
395User Manual 1335.5337.02 ─ 05
larly since faulty commands from the controller to the instrument are recorded there as
well.
16.5.5 Reset values of the status reporting system
The following table contains the different commands and events causing the status
reporting system to be reset. None of the commands, except *RST and
SYSTem:PRESet, influence the functional instrument settings. In particular, DCL does
not change the instrument settings.
Table 16-6: Resets of the status reporting system
Event
DCL, SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYSTem:
PRESet
STATus:
PRESet
*CLS
Clear STB, ESR - - - yes
Clear SRE, ESE - - - -
Clear EVENt parts of the registers - - - yes
Clear ENABle parts of all OPERation and
QUEStionable registers;
Fill ENABle parts of all other registers
with "1".
- - yes -
Fill PTRansition parts with "1";
Clear NTRansition parts
- - yes -
Clear error queue - - - yes
Clear output buffer yes 1) 1) 1)
Clear command processing and input
buffer
yes - - -
1) The first command in a command line that immediately follows a <PROGRAM MESSAGE TERMINA-
TOR> clears the output buffer.