Acquisition and waveform setup
R&S
®
MXO 4 Series
107User Manual 1335.5337.02 ─ 05
"sin (x)/x"
Two adjacent ADC sample points are connected by a sin(x)/x curve,
and also the adjoining sample points are considered by this curve.
The interpolated points are located on the resulting curve. This inter-
polation method is precise and shows the best signal curve.
"Sample/Hold"
The ADC sample points are displayed like a histogram. For each
sample interval, the voltage is taken from the sample point and con-
sidered as constant, and the intervals are connected with vertical
lines. Thus, you see the discrete values of the ADC - the measured
samples.
Remote command:
ACQuire:INTerpolate on page 437
6.2.3 Segmentation settings
Access: "Menu" > "Acquisition" > "Segmented" tab.
You can acquire a limited number of segments, or the maximum number.
Acquire maximum
The setting takes effect for N-single acquisitions ([Single]). During continuous acquisi-
tion ([Run / Stop]), the history is used completely.
If "On", "Segments (N-Single)" is set to the maximum number of available segments,
which is shown in "Max available". Thus, the history is filled completely with an N-sin-
gle acquisition. "Segments (N-Single)" is read-only, you cannot change the value.
If "Off", set the number of segments in "Segments (N-Single)", which is the same set-
ting as "N-single/Avg count" on the "Setup" tab. Thus you can acquire a fast segmenta-
tion acquisition series and history of defined length with [Single].
You can stop the running acquisition before the series is completed.
Acquisition