Remote control commands
R&S
®
MXO 4 Series
590User Manual 1335.5337.02 ─ 05
Manual operation: See "Show decode table" on page 278
SBUS<sb>:FORMat <DataFormat>
Sets the number format for decoded data values of the indicated serial bus. It defines
the format in the decode table, and in the combs of the decoded signal on the screen.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<DataFormat> HEX | OCT | BIN | ASCII | ASCii | SIGN | USIG
*RST: HEX
Usage: Asynchronous command
Manual operation: See "Data format" on page 279
SBUS<m>:ZCOupling <ZoomCoupling>
If enabled, the protocol decode zoom and result table are synchronized.
Suffix:
<m>
.
1...4, index of the serial bus
Parameters:
<ZoomCoupling> ON | OFF
*RST: OFF
Manual operation: See "Zoom coupling" on page 279
SBUS<m>:THReshold <ShwThresLines>
If enabled, the threshold lines are displayed in the diagram.
The lable of the threshold line is set according to the signal it is reffering to.
Suffix:
<m>
.
1..4
Parameters:
<ShwThresLines> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Show threshold lines" on page 277
SBUS<m>:SETReflevels
Executes the measurement of reference levels and sets the thresholds to the middle
reference level of the measured amplitude.