Remote control commands
R&S
®
MXO 4 Series
707User Manual 1335.5337.02 ─ 05
Suffix:
<wg>
.
1..2
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 366
WGENerator<wg>:MODulation:AM:DEPTh
<Depth>
Sets the modulation depth, the percentage of the amplitude range that is used for AM
modulation.
Suffix:
<wg>
.
1..2
Parameters:
<Depth> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Depth" on page 366
WGENerator<wg>:MODulation:AM:FREQuency <Frequency>
Sets the frequency of the modulation waveform for AM frequency modulation.
Suffix:
<wg>
.
1..2
Parameters:
<Frequency> Range: 0.001 to 1000000
Increment: 1
*RST: 1000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Frequency" on page 366
WGENerator<wg>:MODulation:AM:SYMMetry <RampSymmetry>
Sets the symmetry for the AM ramp modulation waveform, the percentage of time that
the waveform is rising.
Waveform generator (option R&S
MXO4-B6)