Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 477

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
477User Manual 1335.5337.02 ─ 05
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Clock edge" on page 154
TRIGger:EVENt<ev>:SETHold:CSOurce:LEVel
<ClockLevel>
Sets the voltage level for the clock signal.
Both the clock level and the clock edge define the starting point for calculation of the
setup and hold time.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Clock level" on page 155
TRIGger:EVENt<ev>:SETHold:CSOurce[:VALue] <ClockSource>
Selects the input channel of the clock signal.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<ClockSource> C1 | C2 | C3 | C4
The following values are also accepted:
C1 = CHAN1 = CHANnel1, C2 = CHAN2 = CHANnel2,
C3 = CHAN3 = CHANnel3, C4 = CHAN4 = CHANnel4
*RST: C1
Usage: Asynchronous command
Manual operation: See "Clock source" on page 154
TRIGger:EVENt<ev>:SETHold:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
Trigger
Loading ...
Loading ...
Loading ...