Trigger
R&S
®
MXO 4 Series
137User Manual 1335.5337.02 ─ 05
B-trigger and R-trigger are configured in the same way as the A-trigger. You can con-
figure a delay between the A- and B-trigger, and define a number of fulfilled B-trigger
conditions to be ignored. The reset condition R can be a timeout or a trigger condition,
or a combination of both.
Trigger sequence
Selects the type of the trigger sequence.
"A B R"
Triggers if all conditions of A- and B-events, as well as additional
delay and count, and optional reset timeout and/or R-event conditions
are fulfilled.
Remote command:
TRIGger:MEVents:AEVents on page 459
Delay
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Remote command:
TRIGger:MEVents:SEQuence<se>:DELay on page 460
B event count
Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trig-
ger causes the trigger event. The waiting time for B-triggers can be restricted with a
reset condition: timeout or reset event.