Remote control commands
R&S
®
MXO 4 Series
644User Manual 1335.5337.02 ─ 05
SBUS<sb>:CAN:DATA:THReshold <SIC Threshold>
Sets a user-defined threshold value for the data line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<SIC Threshold>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 320
SBUS<sb>:CAN:FDATa:DBITrate
<FDBitrate>
Sets the bit rate of the data phase.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<FDBitrate> Range: 10000 to 15000000
Increment: 100
*RST: 1000000
Default unit: bps
Usage: Asynchronous command
Manual operation: See "FD bit rate" on page 319
SBUS<sb>:CAN:FDATa:SAMPlepoint <FDSamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<FDSamplePoint> Range: 12 to 96
Increment: 5
*RST: 66
Default unit: %
Usage: Asynchronous command
Manual operation: See "FD sample point" on page 319
SBUS<sb>:CAN:SIC:HYSTeresis <Hysteresis>
Sets a hysteresis value for the SIC transceiver mode.
Suffix:
<sb>
.
1...4, index of the serial bus
Protocols