Measurements
R&S
®
MXO 4 Series
219User Manual 1335.5337.02 ─ 05
Slope ← Delay measurement
Sets the edge of each source, between which the delay is measured: positive, nega-
tive, or either of them.
Remote command:
MEASurement<mg>:AMPTime:DELay<n>:SLOPe on page 533
Direction ← Delay measurement
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Remote command:
MEASurement<mg>:AMPTime:DELay<n>:DIRection on page 532
Delay to trigger measurement settings
Delay to trigger measures the time between the trigger point and the following slope of
a waveform. The delay between the trigger and the slope can be high compared to the
accuracy of the acquisition, and the trigger point can even be outside of the current
acquisition.