Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 232

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Spectrum analysis
R&S
®
MXO 4 Series
232User Manual 1335.5337.02 ─ 05
Full span
Displays the full frequency span.
CF span, Start Stop
Selects if the frequency span is defined through a "Center"/"Span" pair or through the
"Start"/"Stop".
Center
Defines the position of the displayed frequency range, which is (Center - Span/2) to
(Center + Span/2). The width of the range is defined using the frequency span setting.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:CENTer on page 560
Span
The span is specified in Hertz and defines the width of the displayed frequency range,
which is (Center - Span/2) to (Center + Span/2). The position of the span is defined
using the "Center" setting.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:SPAN on page 560
Start
Defines the start frequency of the displayed frequency span.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:STARt on page 561
Stop
Sets the stop frequency of the displayed frequency span.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:STOP on page 561
Auto RBW
Couples the frequency span to the "RBW" setting.
If span and RBW values are coupled, changing the span also changes the RBW.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:BANDwidth[:RESolution]:AUTO
on page 559
Span/RBW
Defines the coupling ratio for Span/RBW.
This setting is only available if "Auto RBW" is enabled.
Remote command:
CALCulate:SPECtrum<sp>:FREQuency:BANDwidth[:RESolution]:RATio
on page 559
Spectrum setup
Loading ...
Loading ...