Measurements
R&S
®
MXO 4 Series
201User Manual 1335.5337.02 ─ 05
Vertical cursor 1, Vertical cursor 2
Defines a label to be displayed with the vertical cursors.
By default, the cursors are labeled as Cu1.1, Cu1.2, Cu2.1, Cu2.2.
Horizontal cursor 1, Horizontal cursor 2
Defines a label to be displayed with the horizontal cursors.
9.1.3.3 Peak search tab
Access: [Cursor] key > "Peak search" tab
The settings on this tab are only available in spectrum mode, i.e. the source of the cur-
sor measurement is a spectrum. In this case, the cursors can indicate the results of a
peak search on the waveform. You can define which peaks the instrument determines
by defining the noise reject settings.
Absolute peak
Sets both cursors to the absolute peak value.
Remote command:
CURSor<cu>:MAXimum[:PEAK] on page 556
Next peak
Cursor 2 is set to the next smaller absolute peak from the current position.
Remote command:
CURSor<cu>:MAXimum:NEXT on page 557