Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 39

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Getting Started
R&S
®
MXO 4 Series
39User Manual 1335.5337.02 ─ 05
Offset moves the vertical center of the selected channel to the offset value.
If the selected waveform is a math or reference waveform, serial bus, or logic channel,
the knob changes its vertical position.
The knob lights up in the color of the selected waveform. Turn clockwise to move up
the waveform. To set the value to zero, press the knob.
The current offset value is shown in the channel icon.
Remote command:
CHANnel<ch>:POSition on page 431
CHANnel<ch>:OFFSet on page 431
[Scale]
Sets the vertical scale in Volts per division. The vertical scale defines the displayed
amplitude of the selected waveform.
The [Scale] knob lights up in the color of the selected waveform.
Turn the knob clockwise to stretch the waveform. Doing so, the scale value V/div
decreases. Press the knob to toggle between fine and coarse adjustment. For analog
waveforms, the scale value is shown in the signal icon.
To get the maximum resolution of the waveform amplitude, make sure that the wave-
forms cover most of the diagram's height.
Remote command:
CHANnel<ch>:SCALe on page 430
CALCulate:MATH<m>:VERTical:SCALe[:VALue] on page 500
REFCurve<rc>:SCALe on page 509
[Gen]
Opens the "Waveform Generator" dialog, if option R&S MXO4-B6 is installed.
The waveform generator can generate various function and arbitrary waveforms,
sweeps, and parallel patterns. For detailed specifications, refer to the data sheet.
[Logic]
Opens the "Logic" dialog to configure and enable the logic buses L1 to L4. A logic bus
(or parallel bus) has up to 16 logic (digital) channels. If logic buses are active, pressing
the key toggles these buses.
The key lights up if at least one logic bus is enabled.
[Ref]
Opens the "Reference" dialog box, where you can configure and display reference
waveforms. Press the key repeatedly to switch the reference waveform.
If a reference waveform is selected, the vertical rotary knobs are illuminated in white or
light gray (default colors), depending on the selected waveform.
[Math]
Opens the "Math" dialog, where you can configure the calculation of mathematical
waveforms using various mathematic operations on other waveforms. Press the key
repeatedly to toggle the selected math waveform. If no math waveform is active, the
key closes the dialog.
Instrument tour
Loading ...
Loading ...
Loading ...