Acquisition and waveform setup
R&S
®
MXO 4 Series
110User Manual 1335.5337.02 ─ 05
The "Speed" dialog shows information on the current acquisition performance values of
the R&S MXO 4.
You can see the number of "Acquisitions per second", "Average blind time" and the
"Percentage of time captured".
6.3 Vertical setup
The controls and parameters of the vertical system adjust the vertical scale and posi-
tion of the waveform, and the waveform display. The probe settings also belong to the
vertical setup.
The signal icons at the bottom of the display show the most important actual settings
for each channel. The signal icon also indicates if the incoming data is clipped before
processing, i.e. the input range of the ADC is exceeded.
There are several ways to adjust vertical settings:
Use the keys and knobs in the Vertical functional block of the front panel to select
the channel, to scale the waveform, and to set the position or offset.
See Chapter 3.2.4.4, "Vertical controls", on page 38.
Drag one finger vertically on the screen to change the position of the selected
channel waveform.
Spread or pinch two fingers in vertical direction to change the vertical scale of the
selected waveform.
Use the "Vertical" dialog to adjust all vertical settings. See:
Chapter 6.3.2, "Vertical Setup settings", on page 111
Chapter 6.3.3, "Bandwidth settings", on page 114
Chapter 6.4, "Probes", on page 117
Chapter 6.3.5, "Other vertical settings", on page 115
Vertical setup