Remote control commands
R&S
®
MXO 4 Series
451User Manual 1335.5337.02 ─ 05
PROBe<ch>:SETup:ADVanced:FILTer <State>
Activates the lowpass filter in the probe control box. The filter frequency depends on
the probe type and is indicated on the probe control box.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Bandwidth limit" on page 127
PROBe<ch>:SETup:ADVanced:RANGe
<ProbeRange>
Sets the voltage range of an R&S RT-ZHD probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<ProbeRange> AUTO | MHIGh | MLOW
AUTO
The voltage range is set with CHANnel<ch>:SCALe.
MHIGh
Sets the higher voltage range of the connected probe. To query
the value, use PROBe<ch>:SETup:ATTenuation[:AUTO]?.
MLOW
Sets the lower voltage range of the connected probe. To query
the value, use PROBe<ch>:SETup:ATTenuation[:AUTO]?.
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "Range" on page 127
PROBe<ch>:SETup:ADVanced:PMToffset
Sets the measured ProbeMeter value as offset. Thus, the value is considered in mea-
surements.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Usage: Setting only
Asynchronous command
Manual operation: See "ProbeMeter to offset" on page 126