Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 283

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Protocol analysis
R&S
®
MXO 4 Series
283User Manual 1335.5337.02 ─ 05
13.2 SPI bus (option R&S MXO4-K510)
The Serial Peripheral Interface SPI is used for communication with slow peripheral
devices, in particular, for transmission of data streams.
13.2.1 The SPI protocol
A 4-channel instrument is required for full support of the SPI protocol, or the MSO
option R&S MXO4-B1.
The Serial Peripheral Interface SPI is used for communication with slow peripheral
devices, in particular, for transmission of data streams.
Main characteristics of SPI are:
Master-slave communication
No device addressing; The slave is accessed by a chip select, or slave select line.
No acknowledgement mechanism to confirm receipt of data
Duplex capability
Most SPI buses have four lines, two data and two control lines:
Clock line to all slaves (SCLK)
Slave Select or Chip Select line (SS or CS)
Master data output, slave data input (MOSI or SDI)
Master data input, slave data output (MISO or SDO)
When the master generates a clock and selects a slave device, data may be transfer-
red in either or both directions simultaneously.
Figure 13-2: Simple configuration of SPI bus
The data bits of a message are grouped by following criteria:
A word contains a number of successive bits. The word length is defined in the pro-
tocol configuration.
A frame contains a number of successive words, at least one word.
13.2.2 SPI configuration
13.2.2.1 Setup
Access: [Apps] key > "Protocol" tab > "SPI" > "Setup".
SPI bus (option R&S
MXO4-K510)
Loading ...
Loading ...
Loading ...