Parameters:
<Range> Range: 0.01 to 10
Increment: 0.01
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
CHANnel<ch>:OFFSet <Offset>
Sets the offset voltage, which corrects an offset-affected signal. The vertical center of
the selected channel is shifted by the offset value and the signal is repositioned within
the diagram.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Offset> Range: -1 to 1
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Position]" on page 38
CHANnel<ch>:POSition <Position>
Moves the selected signal up or down in the diagram. While the offset sets a voltage,
position is a graphical setting given in divisions. The visual effect is the same as for off-
set.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Position> Positive values move up the waveform, negative values move it
down.
Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
Manual operation: See "[Position]" on page 38
CHANnel<ch>:COUPling <Value>
Sets the connection of the channel signal, i.e. the input impedance (coupling) and a fil-
ter (termination). The command determines what part of the signal is used for wave-
form analysis and triggering.