Getting Started
R&S
®
MXO 4 Series
43User Manual 1335.5337.02 ─ 05
d) Tap "Start Alignment". Wait until the alignment has finished successfully.
2.
Connect the probe's ground connector to the ground pin , and the tip to the
square wave pin .
3. Press the [Preset] key.
4. Connect the probe to the input connector [C1].
The instrument recognizes the probe, and a signal is displayed in the diagram.
5. Tap the signal icon of C1 to open the vertical settings.
6. In the "Vertical" > "Setup" dialog, adjust the following:
"Vertical scale" = 600 mV/div
"Offset" = 900 mV
7. Turn the horizontal [Scale] knob to set the horizontal scale to 1 ms/div.
8. Press the [Source] key to set the trigger source to C1. The color of the key indi-
cates the selected channel.
9. Turn the [Level] knob to set the trigger level to 300 mV.
10. Check the rectangle signal on the screen. The displayed signal should have an
amplitude of about 3.3 V, which covers 5.5 divisions.
Instrument tour