Remote control commands
R&S
®
MXO 4 Series
501User Manual 1335.5337.02 ─ 05
CALCulate:MATH<m>:VERTical:SCALe:MODE <VertSclMd>
Sets how the vertical scale is adapted to the current measurement results. By default,
scaling is done automatically to provide an optimal display. However, if necessary, you
can define scaling values manually to suit your requirements.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<VertSclMd> MANual | AUTO
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "Scale mode" on page 178
CALCulate:MATH<m>:DATA:HEADer?
Returns the header of math waveform data. The header contains attributes of the
waveform.
Table 17-3: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval. For most
waveforms, the result is 1. For envelope waveforms,
it is 2. If the number is 2, the number of returned
values is twice the number of samples (record
length).
1
Suffix:
<m>
.
1...5, index of the math waveform
Example:
CALC:MATH1:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1
Usage: Query only
Asynchronous command
CALCulate:MATH<m>:DATA:STYPe?
Returns the signal type of the source of the math waveform.
Suffix:
<m>
.
1...5, index of the math waveform
Return values:
<SignalType> SOUR | CORR | MEAS | NONE
SOURce = normal signal
Waveform analysis