Remote control commands
R&S
®
MXO 4 Series
479User Manual 1335.5337.02 ─ 05
Example:
TRIG:EVEN1:SOUR C1
TRIG:EVEN1:SLOP POS
TRIG:EVEN1:STAT:QUAL:ANAL:CHAN2:HLX HIGH
TRIG:EVEN1:STAT:QUAL:ANAL:CHAN3:HLX LOW
TRIG:EVEN1:STAT:QUAL:ANAL:CHAN4:HLX HIGH
TRIG:EVEN1:STAT:QUAL:LOG AND
Usage: Asynchronous command
Manual operation: See "Source: channel states" on page 159
TRIGger:EVENt<ev>:STATe:QUALify:LOGic <StateOperator>
Defines the logic combination of the channels and their states.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<StateOperator> AND | OR
*RST: AND
Usage: Asynchronous command
Manual operation: See "State Operator" on page 159
TRIGger:EVENt<ev>:STATe:SLOPe <Slope>
Sets the edge direction for the trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Slope" on page 156
TRIGger:EVENt<ev>:STATe:QUALify:DIGital:LOGic <Logic>
Defines the logic combination of the channels and their states.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Logic> AND | OR
*RST: AND
Usage: Asynchronous command
Manual operation: See "State Operator" on page 159