Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 435

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
435User Manual 1335.5337.02 ─ 05
Position Meaning Example
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval. For most
waveforms, the result is 1. For peak detect and
envelope waveforms, it is 2. If the number is 2, the
number of returned values is twice the number of
samples (record length).
1
Suffix:
<ch>
.
1 to 4, index of the analog channel
<wf> 1, can be omitted
Example:
CHAN1:WAV1:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1
Usage: Query only
Asynchronous command
CHANnel<ch>[:WAVeform<wf>]:DATA[:VALues]? [<Offset>][,<Length>]
Returns the data of the channel waveform points for transmission from the instrument
to the controlling computer. The data can be used in MATLAB, for example.
Without parameters, the complete waveform is retrieved. Using the offset and length
parameters, data can be retrieved in smaller portions, which makes the command
faster. If you send only one parameter, it is interpreted as offset, and the data is
retrieved from offset to the end of the waveform.
To set the export format, use FORMat[:DATA] on page 407.
Suffix:
<ch>
.
1 to 4, index of the analog channel
<wf> 1, can be omitted
Query parameters:
[<Offset>] Number of offset waveform points.
Range: 0 to m. Limit: n + m <= record length
[<Length>] Number of waveform points to be retrieved.
Range: 1 to n. Limit: n + m <= record length
Return values:
<Data> List of values according to the format and content settings.
Example: Retrieve the complete channel 1 waveform, only Y-values:
FORM ASC
CHAN1:WAV1:DATA?
<-- -0.125000,-0.123016,-0.123016,-0.123016,-0.123016,-0.123016,...