Remote control commands
R&S
®
MXO 4 Series
478User Manual 1335.5337.02 ─ 05
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<HoldTime> Range: -9.9999E-08 to 1E-07
Increment: 1E-09
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Hold time" on page 155
TRIGger:EVENt<ev>:SETHold:STIMe
<SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<SetupTime> Range: -9.9999E-08 to 1E-07
Increment: 1E-09
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Setup time" on page 155
17.9.12 State trigger
TRIGger:EVENt<ev>:STATe:QUALify:ANALog:CHAN<n>:HLX........................................... 478
TRIGger:EVENt<ev>:STATe:QUALify:LOGic.....................................................................479
TRIGger:EVENt<ev>:STATe:SLOPe.................................................................................479
TRIGger:EVENt<ev>:STATe:QUALify:DIGital:LOGic.......................................................... 479
TRIGger:EVENt<ev>:STATe:QUALify:DIGital:CHAN<n>:HLX............................................. 480
TRIGger:EVENt<ev>:STATe:QUALify:ANALog:CHAN<n>:HLX <HLX>
Set the state for each channel. For the state trigger, the clock source is indicated and
does not get a state.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<HLX> HIGH | LOW | DONTcare
State of the individual channels
*RST: DONTcare
Trigger
Loading ...
Loading ...
Loading ...