Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 511

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
511User Manual 1335.5337.02 ─ 05
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Enable" on page 191
REFCurve<rc>:RESCale:VERTical:FACTor <ScaleFactor>
Sets the vertical scale factor. A factor greater than 1 stretches the waveform vertically,
a factor lower than 1 compresses the curve.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<ScaleFactor> Range: -1000000 to 1000000
Increment: 0.1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Scale factor" on page 191
REFCurve<rc>:RESCale:VERTical:OFFSet <Offset>
The vertical offset moves the reference waveform vertically. Enter a value with the unit
of the waveform.
Like vertical offset of a channel waveform, the offset of a reference waveform is sub-
tracted from the measured value. Negative values shift the waveform up, positive val-
ues shift it down.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<Offset> Range: -1E+26 to 1E+26
Increment: 1E-06
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Vertical offset" on page 191
REFCurve<rc>:RESCale:VERTical:STATe <State>
If enabled, the vertical offset and factor are applied to the reference waveform.
Stretching and offset change the display of the waveform independent of the vertical
scale and position.
Waveform analysis