Remote control commands
R&S
®
MXO 4 Series
560User Manual 1335.5337.02 ─ 05
CALCulate:SPECtrum<sp>:FREQuency:CENTer <Center>
Defines the position of the displayed frequency range, which is (Center - Span/2) to
(Center + Span/2). The width of the range is defined using the frequency span setting.
Suffix:
<sp>
.
1
Parameters:
<Center> Range: -1E+26 to 1E+26
Increment: 0.01
*RST: 0
Usage: Asynchronous command
Manual operation: See "[Position]" on page 37
CALCulate:SPECtrum<sp>:FREQuency:SCALe
<XAxisMode>
Defines the scaling method for the frequency axis (x-axis) of the spectrogram.
Suffix:
<sp>
.
1
Parameters:
<XAxisMode> LIN | LOG
LIN: linear scaling
LOG: logarithmic scaling
*RST: LIN
Usage: Asynchronous command
Manual operation: See "Frequency axis" on page 235
CALCulate:SPECtrum<sp>:FREQuency:SPAN <Span>
The span is specified in Hertz and defines the width of the displayed frequency range,
which is (Center - Span/2) to (Center + Span/2). The position of the span is defined
using the "Center" setting.
Suffix:
<sp>
.
1
Parameters:
<Span> Range: 0 to 1E+26
Increment: 0.01
*RST: 0.01
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 37