Remote control commands
R&S
®
MXO 4 Series
405User Manual 1335.5337.02 ─ 05
Manual operation: See "Instrument" on page 68
*OPC
Operation complete
Sets bit 0 in the event status register when all preceding commands have been execu-
ted. This bit can be used to initiate a service request. The query writes a "1" into the
output buffer when all preceding commands have been executed, which is useful for
command synchronization.
*OPT?
Option identification query
Queries the options included in the instrument. For a list of all available options and
their description, refer to the data sheet.
Return values:
<Options> The query returns a list of options. The options are returned at
fixed positions in a comma-separated string. A zero is returned
for options that are not installed.
Usage: Query only
*RCL <Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
*SAV with the associated number.
The stored instrument settings do not include waveform generator settings.
It also activates the instrument settings which are stored in a file and loaded using
MMEMory:LOAD:STATe.
*RST
Reset
Sets the instrument to a default status. The default settings are indicated in the
description of commands. Default settings are fixed, they are not changed by user-
defined preset.
The command does not affect the waveform generator settings. To reset the generator,
use WGENerator<wg>:PRESet.
Usage:
Setting only
Common commands