Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 341

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Protocol analysis
R&S
®
MXO 4 Series
341User Manual 1335.5337.02 ─ 05
"Identifier +
Data"
Sets the trigger to a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
The identifier conditions are the same as for the "Identifier" trigger
type, see Identifier setup: Condition, Frame ID min, Frame ID max.
Data conditions are set with Data setup: Condition, Pattern, Position.
"Wakeup
frame"
Triggers after a wakeup frame.
"Error condi-
tion"
Identifies various errors in the frame, see Error conditions.
Remote command:
TRIGger:LIN:TYPE on page 675
Identifier setup: Condition, Frame ID min, Frame ID max
The identifier setup consists of the condition and one or two identifier pattern.
Condition
Sets the comparison condition to a specific value or a range.
"Frame ID min"
Defines the bit pattern of the identifier. Enter only the 6-bit identifier
without parity bits, not the protected identifier.
"Frame ID
max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger:LIN:IMIN on page 677
TRIGger:LIN:IMAX on page 677
TRIGger:LIN:ICONdition on page 677
Data setup: Condition, Pattern, Position
The data setup consists of the transfer direction, the number of bytes, the condition,
and one data pattern.
Condition
Sets the operator to define a specific data pattern ("Equal" or "Not
equal") or a range smaller or bigger than a specific pattern.
"Pattern"
Defines the data pattern.
"Position"
Sets the number of data events that are ignored, before trigger condi-
tion check of the data starts.
Remote command:
TRIGger:LIN:DCONdition on page 676
TRIGger:LIN:DMIN on page 676
TRIGger:LIN:DPOSition on page 676
LIN (option R&S
MXO4-K520)
Loading ...
Loading ...
Loading ...